About
You can change the volume with the Volume Change Tool Kit and use it as your original volume pipette.
[Recommended volume range]
It reaches the highest precision at the preset volume. Select a model which has volume bigger than, and as close as possible to the volume you intend to customize to.
| P / No. |
Recommended volume range |
| NT-F10 | 1µL~10μL |
| NT-F20 | 2µL~20μL |
| NT-F100 | 10µL~100μL |
| NT-F200 | 20µL~200μL |
| NT-F1000 | 100µL~1000μL |
| NT-F5000 | 500µL~5000μL |
■容量変更の手順

① Insert a coin or other thin hard object beneath the fixation cap under the push button and remove the fixation cap gradually.

② Pull off the eject cone and take off the eject button and the spring from the top of the body.

③ Fit the lock spanner on the pipette lock and rotate it unticlockwise to unlock.

④ Adjust the volume by revolving the push button referring to the notch at the back of the button and lock it again.

⑤ Fit a tip on and pre-wet it with distilled water. Then pipette a distilled water into the container placed on the balance.
⑥ Repeat 3,4,5 in order to make adjustments and lock it when the setting has reached the target volume.

⑦ Fit on eject button, spring, eject cone and fixation cap in reverse order of disassembly.
*Push the fixation button strongly with both of your hands.